>(3) In the Emergency treatment of anaphylactic reactions Guidelines for healthcare providers the preferred needle length is 25 mm for adrenaline injectors to access muscle in most people. I heard during expert evidence that Epipen needle length was 16mm - suitable according to the UK Resuscitation Council for “pre-term or very small infants”. The use of needles which access only subcutaneous tissue and not muscle is in my view inherently unsafe. An alternative autoinjector, Emerade has a 24 mm needle.
>(4) The dose of adrenaline in Epipen is 300mcg. The UK Resuscitation Council recommends a standard emergency dose of 500mcg. Emerade contains a dose including 500mcg. The combination of what my expert told me was an inadequate dose of adrenaline for anaphylaxis and an inadequate length needle raises serious safety concerns.
And yet we still have stuff like this: https://www.judiciary.uk/publications/natasha-ednan-laperous...
>(3) In the Emergency treatment of anaphylactic reactions Guidelines for healthcare providers the preferred needle length is 25 mm for adrenaline injectors to access muscle in most people. I heard during expert evidence that Epipen needle length was 16mm - suitable according to the UK Resuscitation Council for “pre-term or very small infants”. The use of needles which access only subcutaneous tissue and not muscle is in my view inherently unsafe. An alternative autoinjector, Emerade has a 24 mm needle.
>(4) The dose of adrenaline in Epipen is 300mcg. The UK Resuscitation Council recommends a standard emergency dose of 500mcg. Emerade contains a dose including 500mcg. The combination of what my expert told me was an inadequate dose of adrenaline for anaphylaxis and an inadequate length needle raises serious safety concerns.